This was not an industry supported study. Funding for this study was provided by Instituto de Salud Carlos III (Fondo de Investigaciones Sanitarias, Ministerio de Sanidad y Consumo) PI050402, Spanish Respiratory Foundation 2005 (FEPAR) and Departamento de Sanidad del Gobierno Vasco (2005111010) and Caja Vital (2005). The authors have indicated no financial conflicts of interest.

Análisis de autorías institucional

Gonzalez Mangado, NicolasAutor o CoautorGarcia Rio, FranciscoAutor o Coautor
Compartir
Publicaciones
>
Artículo

Effectiveness of Home Single-Channel Nasal Pressure for Sleep Apnea Diagnosis

Publicado en:SLEEP. 37 (12): 1953-U115 - 2014-12-01 37(12), DOI: 10.5665/sleep.4248

Autores: Masa, Juan F.; Duran-Cantolla, Joaquin; Capote, Francisco; Cabello, Marta; Abad, Jorge; Garcia-Rio, Francisco; Ferrer, Antoni; Mayos, Merche; Gonzalez-Mangado, Nicolas; de la Pena, Monica; Aizpuru, Felipe; Barbe, Ferran; Montserrat, Jose M.;Spanish Sleep Network

Afiliaciones

Alava University Hospital, Vitoria, Spain. - Autor o Coautor
Alava University Hospital, Vitoria, Spain: Department of Medicine of Basque Country University, Vitoria, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Alava University Hospital, Vitoria, Spain: Sleep Unit, Respiratory Department, Alava University Hospital, Vitoria, Spain. - Autor o Coautor
Alava University Hospital, Vitoria, Spain: Sleep Unit, Respiratory Department, Alava University Hospital, Vitoria, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Bio-Araba Research Institute - Autor o Coautor
Centro de Medicina Respiratoria, Paraná, Argentina. - Autor o Coautor
Clinic Hospital, Barcelona, Spain. - Autor o Coautor
Clinic Hospital, Barcelona, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Clínica Angloamericana, Lima, Perú - Autor o Coautor
Germans Trials i Pujol Hospital, Barcelona, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
IIS-Fundacion Jimenez Diaz, Madrid, Spain. - Autor o Coautor
IIS-Fundación Jimenez Diaz, Madrid, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Infanta Sofía Hospital, San Sebastián de los Reyes, Madrid, Spain. - Autor o Coautor
IRBLleida, Lleida, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
La Paz Hospital, IdiPAZ, Madrid, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Sabadell Hospital, Corporació Sanitària Parc Taulí, Institut Universitari Parc Tauli-UAB, Sabadell, Spain. - Autor o Coautor
Sabadell Hospital, Corporació Sanitària Parc Taulí, Institut Universitari Parc Tauli-UAB, Sabadell, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
San Pedro de Alcantara Hospital, Caceres, Spain and CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
San Pedro de Alcantara Hospital, Caceres, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Son Espases University Hospital, Palma de Mallorca, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Sta Creu i Sant Pau Hospital, Barcelona, Spain: CIBER de enfermedades respiratorias (CIBERES), Madrid, Spain. - Autor o Coautor
Valdecilla Hospital, Santander, Spain. - Autor o Coautor
Virgen del Rocio Hospital, Sevilla, Spain. - Autor o Coautor
‎ Alava Univ Hosp, Resp Dept, Sleep Unit, Vitoria, Spain - Autor o Coautor
‎ Alava Univ Hosp, Vitoria, Spain - Autor o Coautor
‎ BioAraba Res Inst, Vitoria, Spain - Autor o Coautor
‎ CIBER Enfermedades Resp, Madrid, Spain - Autor o Coautor
‎ Clin Angloamer, Lima, Peru - Autor o Coautor
‎ Ctr Med Resp, Parana, Argentina - Autor o Coautor
‎ Germans Trials & Pujol Hosp, Barcelona, Spain - Autor o Coautor
‎ Hosp Clin Barcelona, Barcelona, Spain - Autor o Coautor
‎ Hosp La Paz, IdiPAZ, Madrid, Spain - Autor o Coautor
‎ IIS Fdn Jimenez Diaz, Madrid, Spain - Autor o Coautor
‎ Infanta Leonor Hosp, Madrid, Spain - Autor o Coautor
‎ Infanta Sofia Hosp, Madrid, Spain - Autor o Coautor
‎ Inst Univ Parc Tauli UAB, Sabadell Hosp, Corp Sanitaria Parc Tauli, Sabadell, Spain - Autor o Coautor
‎ IRBLleida, Lleida, Spain - Autor o Coautor
‎ San Pedro de Alcantara Hosp, Caceres, Spain - Autor o Coautor
‎ Son Espases Univ Hosp, Palma de Mallorca, Spain - Autor o Coautor
‎ Sta Creu & Sant Pau Hosp, Barcelona, Spain - Autor o Coautor
‎ Univ Basque Country, Dept Med, Vitoria, Spain - Autor o Coautor
‎ Valdecilla Hosp, Santander, Spain - Autor o Coautor
‎ Virgen del Rocio Hosp, Seville, Spain - Autor o Coautor
Ver más

Resumen

Introduction: Home single-channel nasal pressure (HNP) may be an alternative to polysomnography (PSG) for obstructive sleep apnea (OSA) diagnosis, but no cost studies have yet been carried out. Automatic scoring is simpler but generally less effective than manual scoring. Objectives: To determine the diagnostic efficacy and cost of both scorings (automatic and manual) compared with PSG, taking as a polysomnographic OSA diagnosis several apnea-hypopnea index (AHI) cutoff points. Methods: We included suspected OSA patients in a multicenter study. They were randomized to home and hospital protocols. We constructed receiver operating characteristic (ROC) curves for both scorings. Diagnostic efficacy was explored for several HNP AHI cutoff points, and costs were calculated for equally effective alternatives. Results: Of 787 randomized patients, 752 underwent HNP. Manual scoring produced better ROC curves than automatic for AHI < 15; similar curves were obtained for AHI >= 15. A valid HNP with manual scoring would determine the presence of OSA (or otherwise) in 90% of patients with a polysomnographic AHI >= 5 cutoff point, in 74% of patients with a polysomnographic AHI >= 10 cutoff point, and in 61% of patients with a polysomnographic AHI >= 15 cutoff point. In the same way, a valid HNP with automatic scoring would determine the presence of OSA (or otherwise) in 73% of patients with a polysomnographic AHI >= 5 cutoff point, in 64% of patients with a polysomnographic AHI >= 10 cutoff point, and in 57% of patients with a polysomnographic AHI >= 15 cutoff point. The costs of either HNP approaches were 40% to 70% lower than those of PSG at the same level of diagnostic efficacy. Manual HNP had the lowest cost for low polysomnographic AHI levels (>= 5 and >= 10), and manual and automatic scorings had similar costs for higher polysomnographic cutoff points (AHI >= 15) of diagnosis. Conclusion: Home single-channel nasal pressure (HNP) is a cheaper alternative than polysomnography for obstructive sleep apnea diagnosis. HNP with manual scoring seems to have better diagnostic accuracy and a lower cost than automatic scoring for patients with low apnea-hypopnea index (AHI) levels, although automatic scoring has similar diagnostic accuracy and cost as manual scoring for intermediate and high AHI levels. Therefore, automatic scoring can be appropriately used, although diagnostic efficacy could improve if we carried out manual scoring on patients with AHI < 15.

Palabras clave
apnealinkcost-effectivenessportable monitorApnealinkCost-effectivenessDeviceEventsFlow channelGuidelinesHypopneaMonitorPortable monitorPositive airway pressureRespiratory polygraphySleep apneaValidation

Indicios de calidad

Impacto bibliométrico. Análisis de la aportación y canal de difusión

El trabajo ha sido publicado en la revista SLEEP debido a la progresión y el buen impacto que ha alcanzado en los últimos años, según la agencia WoS (JCR), se ha convertido en una referencia en su campo. En el año de publicación del trabajo, 2014, se encontraba en la posición 24/192, consiguiendo con ello situarse como revista Q1 (Primer Cuartil), en la categoría Clinical Neurology.

Desde una perspectiva relativa, y atendiendo al indicador del impacto normalizado calculado a partir de las Citas Mundiales proporcionadas por WoS (ESI, Clarivate), arroja un valor para la normalización de citas relativas a la tasa de citación esperada de: 1.19. Esto indica que, de manera comparada con trabajos en la misma disciplina y en el mismo año de publicación, lo ubica como trabajo citado por encima de la media. (fuente consultada: ESI 14 Nov 2024)

Esta información viene reforzada por otros indicadores del mismo tipo, que aunque dinámicos en el tiempo y dependientes del conjunto de citaciones medias mundiales en el momento de su cálculo, coinciden en posicionar en algún momento al trabajo, entre el 50% más citados dentro de su temática:

  • Media Ponderada del Impacto Normalizado de la agencia Scopus: 2.46 (fuente consultada: FECYT Feb 2024)
  • Field Citation Ratio (FCR) de la fuente Dimensions: 11.03 (fuente consultada: Dimensions May 2025)

De manera concreta y atendiendo a las diferentes agencias de indexación, el trabajo ha acumulado, hasta la fecha 2025-05-11, el siguiente número de citas:

  • WoS: 36
  • Scopus: 42
  • Europe PMC: 21
  • OpenCitations: 41
Impacto y visibilidad social

Desde la dimensión de Influencia o adopción social, y tomando como base las métricas asociadas a las menciones e interacciones proporcionadas por agencias especializadas en el cálculo de las denominadas “Métricas Alternativas o Sociales”, podemos destacar a fecha 2025-05-11:

  • El uso, desde el ámbito académico evidenciado por el indicador de la agencia Altmetric referido como agregaciones realizadas por el gestor bibliográfico personal Mendeley, nos da un total de: 87.
  • La utilización de esta aportación en marcadores, bifurcaciones de código, añadidos a listas de favoritos para una lectura recurrente, así como visualizaciones generales, indica que alguien está usando la publicación como base de su trabajo actual. Esto puede ser un indicador destacado de futuras citas más formales y académicas. Tal afirmación es avalada por el resultado del indicador “Capture” que arroja un total de: 87 (PlumX).

Con una intencionalidad más de divulgación y orientada a audiencias más generales podemos observar otras puntuaciones más globales como:

  • El Score total de Altmetric: 4.
  • El número de menciones en la red social X (antes Twitter): 2 (Altmetric).

Es fundamental presentar evidencias que respalden la plena alineación con los principios y directrices institucionales en torno a la Ciencia Abierta y la Conservación y Difusión del Patrimonio Intelectual. Un claro ejemplo de ello es:

  • El trabajo se ha enviado a una revista cuya política editorial permite la publicación en abierto Open Access.
Análisis de liderazgo de los autores institucionales

Este trabajo se ha realizado con colaboración internacional, concretamente con investigadores de: Argentina; Peru.